Accredited calibration lab for VOLUME, MASS FLOW, PRESSURE GAS and TEMPERATURE

Since 2017 Mega System Srl is also LAT calibration laboratory N°262T, accredited by ACCREDIA, the only national body authorized by the state to carry out accreditation activities.
Following the positive evaluation obtained in July 2017, ACCREDIA issued the Accreditation Certificate to our Center confirming the adequacy of the Calibration Laboratory to issue LAT certificates for Volume and Volume capacity of gas, in October 2018 for Pressure and in July 2022 for Temperature.

What guarantees the ACCREDIA certification?

The ACCREDIA certification guarantees that our laboratory is adequated to UNI CEI EN ISO / IEC 17025 standard and in compliance with the requirements of other mandatory or voluntary standards.

An accredited calibration center has:

  • a company management system in quality assurance
  • the technical competence to execute accredited calibration of sampling instruments and gas and volume measurement of ambient gases
ACCREDIA LOGO
LAT N° 262T

What are we able to calibrate?

Unit of measureInstrument in calibrationMeasuring rangeUncertainty (*)
Volume of gasGas meters for flow rates from 0,25 l/min to 1 l/min

Gas meters for flow rates from 1,1 l/min to 50 l/min

Gas meters for flow rates from 50 l/min to 100 l/min
From 19 to 75 L

150 L

150 L
0,90%

0,60%

0,70%
Volume of gas capacityMass Flow Controller

Mass Flow Controller

Digital volumetric flow meters
From 0,0051 L/min to 0,25 L/min

From 0,25 L/min to 100 L/min

From 0,0051 L/min to 0,25 L/min

From 0,25 L/min to 100 L/min

From 0,0051 L/min to 0,25 L/min

From 0,25 L/min to 100 L/min
0,35%

0,33%

0,45%

0,40%

0,45%

0,40%

PressureAbsolute pressure meter with digital output (in gas medium) From 40000 to 60000 Pa

From 60001 to 80000 Pa

From 80001 to 100000 Pa

From 100001 to 120000 Pa

From 120001 to 140000 Pa
4,1 Pa

4,7 Pa

5,3 Pa

5,7 Pa

6,3 Pa
Relative pressure meter with digital output (in gas medium)Negative relative pression from 0 to (-)2000 Pa

Negative relative pression from (-)2001 to (-)6000 Pa

Positive relative pression from 0 to (+)2000 Pa

Positive relative pression from (+)2001 Pa to (+)6000 Pa
1,1 Pa

1,2 Pa

1,1 Pa

1,2 Pa
TemperatureIndicators with base metal thermocouples

0 °C

From 5 °C to 70 °C

From 70 °C to 160 °C

From 160 °C to 230 °C

From 260 °C to 400 °C

From 400 °C to 500 °C

From 500 °C to 600 °C

From 600 °C to 700 °C

From 700 °C to 800 °C

From 800 °C to 1068 °C

u1 / u2 (**)

0,05 °C / Uris

0,07 °C / Uris

0,08 °C / Uris

0,12 °C / Uris

0,33 °C / Uris

0,46 °C / Uris

0,63 °C / Uris

0,85 °C / Uris

0,88 °C / Uris

0,89 °C / Uris

Indicators with resistance thermometers0 °C

From 5 °C to 70 °C

From 70 °C to 160 °C

From 160 °C to 230 °C

From 260 °C to 400 °C
u1 / u2

0,02 °C / uris

0,05 °C / uris

0,06 °C / uris

0,07 °C / uris

0,32 °C / uris

(*) The measurements uncertainty is expressed at the trust level of 95%
(**) The expanded measurement uncertainty value is obtained by summing the values of the two components in quadrature with the formula (2√u12 + u22) and is expressed with 2 significant digits, where Uris indicates the standard uncertainty due to the resolution of the instrument in calibration expressed in °C

Which instrument we calibrate?

Our calibration lab releases Calibration Certificate for:

  • Gas meters
  • Mass flow controller
  • Mass flow meter
  • Digital volumetric flow meters
  • Capillary gas dividers
  • Gas dividers with mass flow controller
  • Digital absolute pressure meters
  • Digital relative pressure meters
  • Indicators with base metal thermocouples (K type)
  • Indicators with resistance thermometers (PT 100/1000)

Do you need information about calibration and certificates?